Magnetostrictive level gauge maintenance example

The magnetostrictive level gauge is an extremely accurate measuring instrument that can perform continuous level and interface measurement and provide analog signal output for monitoring and control.
1. The magnetostrictive level gauge shows fluctuations during the day, but the night shift is normal.
[Fault check] The night shift voltage will be high, is the Power Supply abnormal? No problems were found in the inspection of the power supply, sensors, and transmitters. Later, I saw that there was a welding machine at work more than 30 meters away from the meter. If you look at the arc together with electric welding, the liquid level display will fluctuate.
[Troubleshooting] After the electric welder tried to move the ground wire of the welding machine, the magnetostrictive level gauge showed normal recovery.
[Maintenance Summary] This example is a fault caused by electromagnetic interference. The welder welds in several places, too lazy to pull the ground to run around, and the electric welding ground wire is clamped on the support frame of the instrument cable tray, which causes interference to the instrument. To solve the problem of instrument interference, please read the article "Measurement and Control Instrumentation Anti-interference Measures" of Changhui Instrument Network.
2, magnetostrictive level gauge, showing fluctuations and then running the maximum, and finally did not show.
[Fault check] Check that the power supply is normal. The transmitter that displays the abnormality has been powered by a separate power supply, but the output signal is still not normal. When it is not displayed, check that the module of the transmitter is damaged.
[Troubleshooting] Replace the module emergency. Contact the magnetostrictive level gauge supplier to help solve the problem.
[Maintenance Summary] This case is a quality problem of the magnetostrictive level gauge because there are many identical faults in the batch of instruments on the site, and the final result is that the transmitter module burns out. No such failure occurred after the product was replaced.
3, magnetostrictive level gauge, over-range alarm, current up to 21mA.
[Fault check] The power supply and transmitter are normal. Check that the float position is too low below the normal zero when the empty can is found.
[Troubleshooting] Add a stainless steel sleeve to the bottom of the float and return to normal after resetting the range.
[Maintenance Summary] In this case, the difference between the float and the normal zero is too large when there is no liquid level. When the float is at the bottom of the measuring rod, the circular magnetic field of the float and the lower end of the magnetostrictive wire in the measuring rod are insensitive, thus appearing. False alarm phenomenon.

Product Description

Power Recliner Replacement Motor Actuator


24v Linear Actuator For Recliner is an actuator that creates to the circular motion of a conventional ecectric motor.6000N Linear Actuator For Recliners,sofa,lift chair,beds are high quality with low noise and over-load protection are installed.Electric Actuator for Electric Chair Control are used in machine tooles and industrial machinery,in computer peripherals such as disk drives and printers,in valves and dampers,and in many other places where linear motion is required. Linear Actuator for Recliner Chairs are quiet and powerful design.Power recliner motor replacement,the thrust up to 6000N in push,and up to 4000N in pull.Elegant and compact design with small installation dimensions.


Features:Basic and economic design,Up to 6000N load capacity,direct mounting on control box possibility,Long lifetime.


Mimimum installation size(distand between the push holes and the two holes before and after the return of the putter)L=S(stroke)+175mm,the distance between the push and the two holes before and after is extended to =2SX(stroke)+175mm.

Standard protection class:IP 54.Noise level 48 dB.


Application:Electric Linear Actuator for recliner will open and close the lift mechanism of power recliner or lift chair.Connects to motor handset hand control via 5 pin style connector,and connects to the power transformer via the stand 2 pin style connector.


Linear Actuators For Smart Furniture

Linear Actuator,Recliner Chair Parts,Dc Sofa Linear Actuator,Linear Actuators For Smart Furniture

ZHEJIANG XINYI INTELLIGENT DRIVING TECHNOLOGY CO.,LTD , https://www.xinyiactuators.com